Most environmental test chambers look alike on a brochure: a temperature range, a humidity range and a volume. The specifications that decide whether a chamber can actually run your test profile sit underneath those headline figures. This guide covers what to specify, in the order that avoids buying the wrong chamber. If you are still sorting out the terminology, start with types of environmental test chambers explained.


Start With the Test Standard, Not the Chamber

The test you need to run defines the chamber you need. Write down the standards and profiles first, then match chambers to them.


Temperature Range Sets the Refrigeration System

The low end of the range matters most, because it decides how the chamber is refrigerated. A single-stage refrigeration system typically reaches around −40 °C. Reaching −60 °C to −70 °C generally needs a cascade (two-stage) system, which costs more and uses more power. Specify the lowest temperature your standard requires, plus a small margin, rather than the lowest you can buy.


Read the Humidity Envelope, Not Just the Humidity Range

A chamber listed as “20–98% RH” cannot reach every humidity at every temperature. Humidity control only works inside a temperature window, and the low-humidity, low-temperature corner is the hardest to hold. Ask for the chamber’s humidity envelope chart and check that every point of your profile sits inside it. Humidity chambers also need a supply of purified or deionised water; plan for this at installation.


Ramp Rate: Ask How It Was Measured

Ramp rates are quoted in °C per minute, but the figure only means something with its conditions attached.

If you need very fast transitions, a rapid temperature change chamber or a thermal shock chamber may suit better. See our guide to thermal shock vs rapid temperature change chambers.


Fluctuation and Uniformity Decide Whether Results Are Valid

Two figures describe how well a chamber holds conditions. Fluctuation is how much the temperature varies over time at one point. Uniformity is how much it varies from point to point across the workspace at the same moment. Your standard may set limits for both. Ask for measured data at the setpoints you will use, and confirm how it was measured; IEC 60068-3-5 and IEC 60068-3-6 describe methods for confirming chamber performance.


Size the Workspace Around the Load


Benchtop, Reach-In or Walk-In

FormatTypical useWatch for
BenchtopComponents, small assemblies, R&DLimited load and heat dissipation
Reach-in (floor-standing)Most product qualification workDoor size, floor loading, power supply
Walk-in or drive-inLarge equipment, vehicles, batteries, many units at onceSite works, installation time, ventilation

Installation and Service in Australia


Specification Checklist

  1. Standards and profiles the chamber must run
  2. Lowest and highest temperature, with margin
  3. Humidity points, checked against the humidity envelope
  4. Ramp rate, stated as average or linear, loaded or empty, over which span
  5. Fluctuation and uniformity limits at your setpoints
  6. Workspace size, device heat load and cable ports
  7. Power, cooling, water and site access
  8. Local service and calibration

Browse our temperature and humidity chambers, thermal shock chambers and the full range of environmental test chambers supplied and supported in Australia by Finch.

Not sure which chamber fits your profile? Send us your test specification and we’ll help you match it.


Frequently Asked Questions

What is the difference between an environmental chamber and a climatic chamber? The terms are usually used interchangeably. Both refer to chambers that control temperature and, often, humidity. “Environmental chamber” is sometimes used more broadly to include corrosion, UV and altitude chambers.

Do I need humidity control? Only if your standard or failure mode involves moisture, such as damp heat tests or condensation-related faults. A temperature-only chamber is simpler and cheaper.

Can one chamber cover temperature cycling and thermal shock? Not usually. Thermal shock moves the device between two zones in seconds, which a single-zone chamber cannot do.