Most laboratory power amplifiers are voltage amplifiers: they reproduce the input waveform as a scaled voltage and let the load decide the current. For resistive loads that is exactly what you want. For coils, transformer cores and magnetic actuators, the quantity that matters is current, and a voltage-mode amplifier does not control it directly. That is why amplifiers such as the Aigtek ATA-3100B offer both constant-voltage (CV) and constant-current (CC) operation.


CV vs CC: What Each Mode Controls

ModeWhat the amplifier holds to the inputWhat varies with the loadBest for
Constant-voltage (CV)Output voltageCurrentResistive loads, piezo and ultrasonic drive, general signal amplification
Constant-current (CC)Output currentVoltageCoils and electromagnets, transformer core excitation, current sensor testing

The magnetic field a coil produces is proportional to the current through it. In CV mode that current depends on the coil's impedance, which rises with frequency (inductive reactance is 2πfL) and drifts as the copper warms up. A frequency sweep in CV mode therefore produces a field that falls off at higher frequencies and creeps as the coil heats. In CC mode the amplifier adjusts its output voltage to keep the current, and so the field, where you set it.


Why Voltage Headroom Still Matters

Constant current only works while the amplifier has enough voltage to push that current through the load. For an inductive load the required voltage is roughly the current times the load's impedance at that frequency, so the higher the frequency or inductance, the more voltage is needed. Once the amplifier reaches its voltage limit, it can no longer hold the current. When sizing an amplifier for coil work, check both figures: the peak current you need, and the voltage the coil will demand at your highest test frequency. The ATA-3100B's 280Vp-p swing and 45Ap continuous current (52Ap pulsed) are intended to cover both.


Matching the Output Mode to the Load

High-current amplifiers often let you trade voltage for current to suit the load impedance. The ATA-3100B's AC output tables are given for three settings:

Using the wrong setting wastes capability. A 1Ω coil on a high-voltage range will hit the current limit long before the voltage is used, while a 16Ω load on a high-current range will run out of voltage.


Reading Pulse, 5-Minute and 1-Hour Ratings

Power amplifier datasheets often give several output figures for the same load. They are not contradictions; they describe how long the amplifier can sustain each level before heat becomes the limit:

At low impedances the gap between them can be large. Into 1Ω in medium-level mode, the ATA-3100B delivers 47Ap in pulses and 39.6Ap for five minutes, but 21Ap continuously for an hour. Size for the duration your test actually runs.


Typical Applications