Cables act as antennas. Below about 80 MHz, most devices are too small to pick up much RF energy directly, but the cables attached to them are long enough to. IEC 61000-4-6 tests a product’s immunity to that energy by injecting RF directly onto its cables, rather than radiating a field around the whole product. This guide explains what the test involves and what equipment a compliant setup needs.


Where Conducted Immunity Fits

IEC 61000-4-6 covers roughly 150 kHz to 80 MHz; some product standards extend it higher. Radiated immunity under IEC 61000-4-3 takes over from 80 MHz. Most product and generic EMC standards call up both, so a lab testing to one usually needs the other.


Test Levels and Modulation

LevelTest voltage (EMF, unmodulated)
11 V
23 V
310 V
XSet by the product standard

The carrier is amplitude-modulated at 80% with a 1 kHz sine wave. Generic standards typically call for 3 V for residential and light-industrial environments and 10 V for industrial ones; your product standard decides.


Choosing a Coupling Method

The disturbance has to reach the cable with the right common-mode impedance, nominally 150 Ω. The standard gives several ways to do this, in order of preference.

MethodHow it couplesBest forTrade-off
Coupling/decoupling network (CDN)Direct connection to the cable conductorsPower lines and cables with a matching CDN typeNeeds a CDN for each cable type; preferred where one exists
EM clampCombined capacitive and inductive couplingCables with no suitable CDNNeeds a CDN or decoupling on other cables
Current injection probe (bulk current injection)Inductive coupling onto the cable bundleMulti-wire cables, harnessesNeeds more amplifier power and a current monitor
Direct injectionOnto the shield of screened cablesScreened cablesApplies only to screens

Sizing the Amplifier

The standard requires an attenuator of at least 6 dB between the amplifier and the coupling device, to keep the source impedance stable. Add the coupling device’s own loss and the 80% modulation peak, and a 10 V test commonly needs an amplifier in the tens of watts. Clamp and current-injection methods need more power than a CDN. Size the amplifier from the coupling devices you will actually use, with margin. Our guide to selecting an amplifier for IEC 61000-4-3 covers the same trade-offs for radiated testing.


Setting the Level and Running the Sweep

  1. Set the level first. The test level is set in a 150 Ω calibration setup without the device connected, and the amplifier settings are recorded across the frequency range.
  2. Lay out the test. The device sits on an insulating support above a ground reference plane, with coupling devices placed 0.1–0.3 m from it.
  3. Sweep. The frequency is stepped in increments of no more than 1% of the previous frequency.
  4. Dwell. At each step the signal is held at least long enough for the device to respond, and never less than 0.5 s.
  5. Monitor and judge. The device’s behaviour is checked against the performance criteria in the product standard.

Equipment Checklist


How Finches Supplies a Conducted Immunity Setup

Conducted RF immunity setups depend on the cables and connectors of the product being tested, so Finches supplies them as custom integrated systems. The CDNs, clamps and calibration fixtures are chosen around your product and test standard, then paired with an RF power amplifier from RFlight or Aigtek. Contact us with your product standard, test level and cable types to have a system specified.

View EMC amplifiers


Frequently Asked Questions

What is the difference between conducted and radiated immunity? Conducted immunity injects RF directly onto cables below about 80 MHz. Radiated immunity applies an RF field to the whole device, usually from 80 MHz upward.

Can I use an EM clamp instead of buying CDNs? The standard prefers CDNs where a suitable one exists. Clamps are intended for cables where a CDN is not practical.

Is conducted immunity the same as conducted emissions? No. Emissions testing measures noise the device puts onto its cables; see our guide to conducted emissions and LISN selection. Immunity testing checks whether the device withstands noise put onto them.