Cables act as antennas. Below about 80 MHz, most devices are too small to pick up much RF energy directly, but the cables attached to them are long enough to. IEC 61000-4-6 tests a product’s immunity to that energy by injecting RF directly onto its cables, rather than radiating a field around the whole product. This guide explains what the test involves and what equipment a compliant setup needs.
Where Conducted Immunity Fits
IEC 61000-4-6 covers roughly 150 kHz to 80 MHz; some product standards extend it higher. Radiated immunity under IEC 61000-4-3 takes over from 80 MHz. Most product and generic EMC standards call up both, so a lab testing to one usually needs the other.
Test Levels and Modulation
| Level | Test voltage (EMF, unmodulated) |
|---|---|
| 1 | 1 V |
| 2 | 3 V |
| 3 | 10 V |
| X | Set by the product standard |
The carrier is amplitude-modulated at 80% with a 1 kHz sine wave. Generic standards typically call for 3 V for residential and light-industrial environments and 10 V for industrial ones; your product standard decides.
Choosing a Coupling Method
The disturbance has to reach the cable with the right common-mode impedance, nominally 150 Ω. The standard gives several ways to do this, in order of preference.
| Method | How it couples | Best for | Trade-off |
|---|---|---|---|
| Coupling/decoupling network (CDN) | Direct connection to the cable conductors | Power lines and cables with a matching CDN type | Needs a CDN for each cable type; preferred where one exists |
| EM clamp | Combined capacitive and inductive coupling | Cables with no suitable CDN | Needs a CDN or decoupling on other cables |
| Current injection probe (bulk current injection) | Inductive coupling onto the cable bundle | Multi-wire cables, harnesses | Needs more amplifier power and a current monitor |
| Direct injection | Onto the shield of screened cables | Screened cables | Applies only to screens |
Sizing the Amplifier
The standard requires an attenuator of at least 6 dB between the amplifier and the coupling device, to keep the source impedance stable. Add the coupling device’s own loss and the 80% modulation peak, and a 10 V test commonly needs an amplifier in the tens of watts. Clamp and current-injection methods need more power than a CDN. Size the amplifier from the coupling devices you will actually use, with margin. Our guide to selecting an amplifier for IEC 61000-4-3 covers the same trade-offs for radiated testing.
Setting the Level and Running the Sweep
- Set the level first. The test level is set in a 150 Ω calibration setup without the device connected, and the amplifier settings are recorded across the frequency range.
- Lay out the test. The device sits on an insulating support above a ground reference plane, with coupling devices placed 0.1–0.3 m from it.
- Sweep. The frequency is stepped in increments of no more than 1% of the previous frequency.
- Dwell. At each step the signal is held at least long enough for the device to respond, and never less than 0.5 s.
- Monitor and judge. The device’s behaviour is checked against the performance criteria in the product standard.
Equipment Checklist
- RF signal generator covering 150 kHz to 80 MHz (or higher if your standard requires), with 1 kHz AM
- Broadband RF power amplifier sized for your test level and coupling method, such as RFlight EMC amplifiers or the Aigtek ATA-8000 RF power amplifier (100 kHz to 25 MHz, depending on model)
- 6 dB (or greater) attenuator
- CDNs for each power and signal cable type, plus EM clamp or current injection probe where no CDN fits (Finches specifies these as part of a custom integrated system)
- 150 Ω to 50 Ω adapters and calibration fixtures
- Ground reference plane and insulating support
- Control software to run and record the sweep
How Finches Supplies a Conducted Immunity Setup
Conducted RF immunity setups depend on the cables and connectors of the product being tested, so Finches supplies them as custom integrated systems. The CDNs, clamps and calibration fixtures are chosen around your product and test standard, then paired with an RF power amplifier from RFlight or Aigtek. Contact us with your product standard, test level and cable types to have a system specified.
Frequently Asked Questions
What is the difference between conducted and radiated immunity? Conducted immunity injects RF directly onto cables below about 80 MHz. Radiated immunity applies an RF field to the whole device, usually from 80 MHz upward.
Can I use an EM clamp instead of buying CDNs? The standard prefers CDNs where a suitable one exists. Clamps are intended for cables where a CDN is not practical.
Is conducted immunity the same as conducted emissions? No. Emissions testing measures noise the device puts onto its cables; see our guide to conducted emissions and LISN selection. Immunity testing checks whether the device withstands noise put onto them.