LISUN LS8815 impulse winding testers detect inter-turn and inter-layer shorts in coils: 100V to 5kV pulses, 100MSa/s sampling, single or 8-channel models.
Enquire About This Product
Turn-to-turn and layer-to-layer insulation faults in a coil often pass a normal resistance or inductance check but cause failures in service. An impulse winding test finds them by charging a small capacitor to high voltage and discharging it into the coil. The coil and capacitor form an RLC circuit whose damped oscillation changes shape when insulation is weak.
The LS8815 series uses low test energy, so the test is non-destructive, and captures the decaying waveform at up to 100MSa/s (200MSa/s available) to compare it with a reference from a good part. Area, differential area, corona and phase difference comparisons detect inter-turn shorts, inter-layer shorts and reduced insulation. Measurement is fast and repeatable, for motors, mains transformers, relays, switching power supply transformers and degaussing coils.
| Specification | LS8815S-8 | LS8815-5KV | LS8815-3KV |
|---|---|---|---|
| Pulse voltage | 100~5000V, 10V step, 5% ±10V | 100~3000V, 10V step, 5% ±10V | |
| Channels | 8 | 1 | |
| Inductance range | >10μH | ||
| Max pulse energy | 0.25J | ||
| Sampling | 100MSa/s~100kSa/s, 8-bit, 6500 points | ||
| Input impedance | 10MΩ | ||
| Speed | up to 15 tests per second | ||
| Pulses | Test 1~32; degauss 0~16 | ||
| Comparison modes | Area, differential area, corona, phase difference | ||
| Repeatability | Area ±1%; differential area ±1% | ||
| Memory | 200 internal, 500 on USB drive | ||
| Interfaces | Handler, RS232C, USB device, USB host, IEEE488 optional | ||
Inter-turn shorts, inter-layer shorts and reduced insulation within the winding, which resistance and inductance checks can miss.
No. It uses low pulse energy (0.25J maximum).