WB2678A Ground Bond Resistance Tester (10A/25A) – LISUN

WB2678A

LISUN WB2678A ground bond tester: 10A or 25A test current, 0-600mΩ (10A) or 0-200mΩ (25A), ±5%, 1-99s timer and limit alarm, for IEC 60335-1 and 60598-1.

Electrical Safety Test Equipment
Grounding Resistance Tester
Bench-top Instrument
IEC 60335-1; IEC 60598-1; IEC 60950-1; IEC 60601-1; GB 4706.1; GB 7000.1; GB 4943.1; GB 9706.1
Enquire About This Product
LISUN WB2678A Ground Bond Resistance Tester (10A/25A)

Overview

Class I equipment relies on its protective earth to keep exposed metal safe if insulation fails, so safety standards require the resistance between the earth pin and accessible metal parts to be measured at a high test current. The WB2678A is a dedicated ground bond (earth continuity) tester for this test.

It passes a test current of 10A or 25A through the earth path to simulate fault conditions, revealing poor connections and high contact resistance that a low-current ohmmeter would miss. Three LED windows show resistance, test current and test time at once, and the tester offers a 1 to 99 second timer, a preset limit with pass/fail alarm, and manual start and stop.

Typical uses include the earth terminal of household appliances to IEC 60335-1, luminaires to IEC 60598-1, IT equipment to IEC 60950-1 and medical devices to IEC 60601-1, along with industrial and R&D lab testing.

Key Specifications

ItemSpecification
Resistance range0~600mΩ at 10A; 0~200mΩ at 25A
Test current10A / 25A
Test time1~99s
Accuracy±5%
DisplayThree LED windows: resistance, current, time

Features

Applications

Frequently asked questions

Which test current should I use?

The product standard sets it. IEC 60598-1 for luminaires uses at least 10A, while IEC 60335-1 for appliances uses 25A or 1.5 times rated current, whichever is higher.

Why not use a multimeter?

A high current reveals loose or corroded connections that pass at milliamp currents but would fail under a real fault.