LISUN TPB-UL935 heating chamber for thermally protected ballasts to UL 935 and IEC 61347-1: 610mm cube, RT+10 to 100°C, ±2°C uniformity, wooden sample rack.
Enquire About This Product
Ballasts with Class P or other thermal protection must switch off before they overheat. UL 935 and IEC 61347-1 verify this by operating the ballast in a heated enclosure and checking that the protector opens at the right temperature and that surface temperatures stay within limits.
The TPB-UL935 provides a stable heated environment from 10°C above room temperature to 100°C, with ±2°C uniformity and ±0.5°C fluctuation. It has stainless steel inner and outer chambers, ultra-fine glass wool insulation, four 300W strip heaters connected in parallel so that one failed heater does not stop the test, and a wooden sample rack that avoids heat conduction through metal. Three switchable measuring points monitor temperature at different positions.
| Item | Specification |
|---|---|
| Internal size | 610 × 610 × 610mm |
| Temperature range | RT+10°C~100°C |
| Uniformity | ≤±2°C |
| Fluctuation | ≤±0.5°C |
| Chamber | Stainless steel inside and out |
| Insulation | Ultra-fine glass wool |
| Heaters | 4 strip heaters, 300 × 40mm, 300W each, in parallel |
| Sample rack | Wood |
| Measuring points | 3, switchable |
Wood does not conduct heat to or from the ballast, so the temperature it sees comes from the air, as the test intends.