LISUN IK01-07-CAL calibrates spring impact hammers to IEC 60068-2-75 Annex B and JJF 1475: 0–2J range (IK01–IK07), ±0.01J repeatability.
Enquire About This Product
Spring impact hammers are used across appliance, luminaire, switch and enclosure standards to apply a known impact energy, from 0.14J (IK01) up to 2J (IK07). Over time the spring and mechanism can drift, so the hammer must be calibrated to show it still delivers the right energy.
The LISUN IK01-07-CAL is a calibration device built to IEC 60068-2-75 Annex B (and the earlier IEC 60068-2-63 Annex A). The hammer is fired at a pendulum, and the energy it transfers is measured from the pendulum's swing. It can calibrate any spring hammer made to IEC 60068-2-75 with an energy of 0 to 2J, including LISUN's IK01-07 hammer and hammers from other makers.
With repeatability of ±0.01J and pendulum energy losses below 0.002J, the device lets a laboratory check its own hammers between external calibrations, which is useful for laboratories working to ISO/IEC 17025.
| Parameter | Value |
|---|---|
| Measuring range | 0–2J |
| Repeatability | ±0.01J |
| Pendulum energy loss | <0.002J |
| Guide groove outside diameter | 51mm |
| Trigger travel | >30mm |
| Counterweights | One copper weight set |
| IK01 | IK02 | IK03 | IK04 | IK05 | IK06 | IK07 |
|---|---|---|---|---|---|---|
| 0.14J | 0.2J | 0.35J | 0.5J | 0.7J | 1J | 2J |
That is set by your quality system, typically once a year or after heavy use or damage. This device lets you check hammers in-house in between.
Yes, any spring hammer made to IEC 60068-2-75 with an energy up to 2J.