LS-RVT Residual Voltage Tester (Peak Power Cut-off) – LISUN

LS-RVT-2000W / LS-RVT-4000W / LS-RVT-5000W-3P

LISUN LS-RVT residual voltage testers cut the supply at the voltage peak and measure what is left on the pins: 5 to 300V, ≥100MΩ, for IEC 60335-1 and 60601-1.

Electrical Safety Test Equipment
Residual Voltage / Safety Tester
Bench-top Instrument
IEC 60335-1:2020 Clause 22.5; IEC 60601-1:2020 Clause 8.4.3; IEC 60598-1:2024; IEC 60065:2014; IEC 60950-1; GB/T 4706.1-2024; GB 9706.1-2020; GB 7000.1-2015; GB 8898-2011; GB 4943.1-2022
Enquire About This Product
LISUN LS-RVT Residual Voltage Tester (Peak Power Cut-off)

Overview

When an appliance is unplugged, capacitors in its input circuit can leave a voltage on the plug pins that could shock someone touching them. Standards such as IEC 60335-1 limit this voltage one second after disconnection, and the worst case occurs when the supply is cut at the peak of the mains waveform.

The LS-RVT testers switch the product off at the voltage peak and then measure the residual voltage through a high-impedance input of at least 100MΩ (1000MΩ optional) and 25pF, simulating the plug being pulled. Tests can be repeated automatically with results recorded and statistics calculated, and an alarm sounds and the supply is cut if the result exceeds the set limit. An optional function measures stored charge up to 99.9μC for IEC 60601-1 Clause 8.4.3. Models differ in output power, from 2000W single phase to 5000W three phase.

Key Specifications

SpecificationLS-RVT-2000WLS-RVT-4000WLS-RVT-5000W-3P
Max output2000W / 10A, single phase4000W / 16A, single phase5000W / 7.5A, three phase
Output voltageAC 0~242VAC 0~380V
Residual voltage range5~300.0V, 0.1V resolution, ±(1% rdg + 1V)
Test time1~99s
Input impedance≥100MΩ (>1000MΩ optional), 25pF
Alarm range5~300.0V
JudgementAudible and visual alarm with supply cut-off
Stored charge (option)0~99.9μC, to IEC 60601-1 Clause 8.4.3
InterfaceRS-485 (optional)

Features

Applications

Frequently asked questions

Why disconnect at the peak?

The residual voltage is highest when the supply is cut at the peak of the waveform, so this gives the worst case the standards require.