SUG1250 PV Module Impulse Voltage Generator – LISUN

SUG1250-30KV / SUG1250-15KV

LISUN SUG1250 impulse voltage generators for PV modules to IEC 61730-2 MST 14: 1.2/50μs impulse up to 15kV or 30kV into 15–150nF module capacitance.

Conducted Immunity Test Systems
Surge / Impulse Generator
Bench-top Instrument
IEC 61730-2:2023 Clause 10.12 MST 14; IEC 61730-1; IEC 60060-1; GB/T 20047.2; GB/T 20047.1; GB/T 16927.1
Enquire About This Product
LISUN SUG1250 PV Module Impulse Voltage Generator

Overview

IEC 61730-2 requires photovoltaic modules to pass an impulse voltage test (MST 14) that checks the module's insulation can withstand lightning-induced overvoltages. Because a PV module is a large capacitive load, the generator must keep a correct 1.2/50μs waveform across a wide range of module capacitance.

The SUG1250 series is designed for exactly that test. It produces the 1.2/50μs impulse into modules from 15nF to 150nF, selected in nine steps, at up to 30kV (SUG1250-30KV) or 15kV (SUG1250-15KV). A built-in voltage probe shows the peak voltage in real time on a large LCD touch screen.

Key Specifications

SpecificationSUG1250-30KVSUG1250-15KV
Test voltage3~30kV ±3%3~15kV ±3%
Waveform1.2μs ±30% / 50μs ±20%
Wave-shaping networkMatched to load capacitance
PolarityPositive, negative, alternating
Interval8~9,999s
Impulse count1~9,999
TriggerManual, automatic, external
Load capacitance15~150nF in 9 steps

Features

Applications

Frequently asked questions

Which voltage do I need?

IEC 61730-2 sets the impulse voltage from the module's maximum system voltage and class. The 30kV model covers higher system voltages; the 15kV model suits lower ratings.

Why does capacitance matter?

A large module capacitance distorts the impulse, so the generator switches its wave-shaping network to suit the load and keep the 1.2/50μs shape.