LISUN
Home
/
Specialised Environmental & Mechanical Test Equipment

LVD-AT Electrodynamic Vibration System (CNAS) – LISUN

LVD-AT01 / LVD-AT10 / LVD-AT30 / LVD-AT60

LISUN LVD-AT electrodynamic shakers, CNAS (ISO 17025) certified: 1 to 60kN sine and random force, up to 6000Hz, 25–76mm stroke, slip table and controller.

Specialised Environmental & Mechanical Test Equipment
Vibration Test
Floor-Standing Vibration System
IEC 60068-2-6; IEC 60598-1:2020 Clause 4.20; ISO 2247; ISO 13355; ASTM D999; ASTM D4728; ASTM D3580; JJG 189-97
Enquire About This Product
LISUN LVD-AT Electrodynamic Vibration System (CNAS)

Overview

The LISUN LVD-AT series are electrodynamic vibration test systems for sine, random and shock testing, with rated force from 1kN to 60kN. Each system carries CNAS certification under the ILAC-MRA ISO/IEC 17025 framework, and LISUN lists it for full-compliance testing to IEC 60068-2-6, IEC 60598-1 Clause 4.20 and the ISO and ASTM package vibration standards. Typical users are the aviation, aerospace, marine, rail and new-energy industries.

How an electrodynamic shaker works

A drive coil sits in the gap of a strong magnetic field. When the power amplifier drives current through the coil, the electromagnetic force moves the coil and the table attached to it, following the signal from the vibration controller. The force is proportional to the current, so the controller can produce sine sweeps, random spectra and shock pulses.

System components

Uses include vibration fatigue testing, finding resonant frequencies, routine environmental vibration tests, measuring the dynamic response of parts for design, and calibrating accelerometers and other vibration sensors.

Key Specifications

ModelLVD-AT01LVD-AT10LVD-AT30LVD-AT60
Rated force, sine / random1kN10kN30kN60kN
Rated force, shock2kN20kN60kN120kN
Max. sample mass100kg300kg500kg800kg
Frequency range5–6000Hz5–4000Hz5–3000Hz5–2700Hz
Rated sine acceleration650m/s²980m/s²980m/s²1127m/s²
Rated random acceleration650m/s²784m/s²980m/s²784m/s²
Displacement (p-p)25mm51mm51mm76mm
Rated velocity1.6m/s2m/s2m/s2m/s
Moving mass1.5kg10kg25kg53kg
Armature diameterØ120mmØ240mmØ340mmØ445mm
Magnetic flux leakage≤1mT≤1mT≤1mT≤1mT
Shaker size (L × W × H)400 × 365 × 470mm930 × 720 × 760mm1180 × 780 × 1090mm1400 × 1010 × 1400mm
Shaker weight175kg900kg1750kg3500kg
Amplifier rating1kVA10kVA30kVA60kVA
System power3kVA21kVA46kVA96kVA

Features

Applications

Frequently asked questions

How do I size a shaker?

Required force is roughly the total moving mass (armature, fixture and sample) multiplied by the peak acceleration. Send us your sample mass and test profile and we will recommend a model.

What is the difference from the LVD-100KG?

The LVD-100KG is a compact pre-compliance table for sine testing up to 600Hz. The LVD-AT systems are full electrodynamic shakers for random and high-frequency testing.

Related Reading