Overview

Accurate lamp, driver and appliance measurements need a clean, stable supply that does not change with the local mains. The LSP-VARC series are high-precision pure sine wave AC sources with very low harmonic distortion (0.3% or less), voltage and frequency stability within 0.1% and 0.05% over 30 minutes, and a near-zero internal resistance design.

A built-in meter shows output voltage, current, power, power factor and frequency on a large LCD, so the source doubles as a power meter. Protection covers short circuit, over-current, over-voltage, over-power and over-temperature. Over RS232/RS485 it works with LISUN's LSG goniophotometers, LPCE-2 integrating sphere systems, LSRF-3 lamp start and flicker system and LS2090 LED driver tester.

The Pst versions add programmable voltage fluctuation to IEC TR 61547-1:2020, IEC 61000-3-3, IEC 61000-4-15 and IEEE 1453, and work with the LSRF-3 to measure the Pst LM(I) flicker immunity of lighting products.

Key Specifications

SpecificationLSP-500VARCLSP-500VARC-PstLSP-1KVARCLSP-1KVARC-Pst
OutputSingle phase, 2~150V / 2~300V
Max current (2~150V)4.2A4.2A8.4A8.4A
Max current (2~300V)2.1A2.1A4.2A4.2A
Frequency45.00~65.00Hz
Frequency stability≤0.05% / 30min
THD≤0.3% (no load or full resistive load)
Voltage stability≤0.1% / 30min
Load effect≤0.1%
Pst flicker programmingNoYesNoYes
Meter rangesVoltage 55V/220V, 0.1V resolution; current 0.2A/2A, 0.001A resolution; power 0.1W resolution
Meter accuracy±(0.4% rdg + 0.2% rng + 1 digit) for V, I and P
InterfaceRS232 / RS485

Features

Applications

Frequently asked questions

What is the Pst version for?

IEC TR 61547-1 tests how much a lamp's light flickers when the supply voltage fluctuates. The Pst models generate the required voltage fluctuations, and with the LSRF-3 they measure Pst LM(I).

500VA or 1kVA?

Choose by the load: the 500VA model gives 2.1A at up to 300V; the 1kVA model gives 4.2A.