LISUN LISN-PVDC8301 symmetric DC artificial network to CISPR 16-1-2 for conducted emissions on the DC side of PV inverters: 150kHz–30MHz, 1500V, 200A.
Enquire About This Product
Photovoltaic inverters can generate ripple currents on their DC side that travel through the cables and PV modules and radiate as a magnetic field, sometimes causing serious interference. Measurements at the inverter's AC terminals do not reveal this, so the DC side needs its own conducted emission test.
The LISN-PVDC8301 is a symmetric DC artificial network designed to CISPR 16-1-2 for that job. Used with a LISUN EMI-9KB or EMI-9KA receiver, it measures disturbance voltage from 150kHz to 30MHz on DC lines up to 1500V and 200A. A mode switch selects unsymmetrical (one conductor to ground), asymmetrical (common mode) or differential mode measurement.
| Item | Specification |
|---|---|
| Frequency range | 0.15~30MHz |
| Max DC current | 200A |
| Max DC voltage | 1500V |
| Common mode impedance | 150 ±30Ω (switch CM, A or B) |
| Differential mode impedance | 150 ±30Ω (switch DM) |
| Phase at EUT terminals | 0 ±40° |
| Insertion loss (EUT to AE) | >20dB |
| Longitudinal conversion loss | >26dB |
| Voltage division factor | typ. 20 ±3dB (10:1) |
| DC resistance (feed shorted, 25℃) | <50mΩ |
| EUT connectors | M12 terminals with banana sockets |
| Size / weight | 42×28×44cm / approx. 33kg |
DC-side ripple currents can radiate through the PV array and are not visible in AC-side measurements, so a DC network is needed to measure them.
LISUN recommends its EMI-9KB or EMI-9KA receiver; any CISPR 16-1-1 receiver covering 150kHz–30MHz can be used.