LISUN LEDLM-80PL LED lumen maintenance system for single LEDs and chips to IES LM-80-21 and TM-21-19: up to 16 samples, L70/L50, Arrhenius life projection.
Enquire About This Product
The LEDLM-80PL measures how the light output of single LEDs and LED chips falls over time, so you can state their lumen maintenance and projected life. It follows IES LM-80-21 for the measurement and IES TM-21-19 for projecting long-term life, and is used for LED R&D and quality inspection.
Up to 16 samples are tested at once (more on request) inside a GDJW or GW series temperature chamber, powered by a DC12010 constant-current/constant-voltage supply, with a TMP-8 tester monitoring sample temperatures. The system records luminous flux, colour and electrical parameters against time and generates lumen maintenance and life projection reports. You can run the full 6,000-hour test, or use the Arrhenius model to accelerate the projection, against L70, L50 or your own criterion. The results support certification programmes such as ENERGY STAR.
For complete LED luminaires rather than single LEDs, see the LEDLM-84PL.
| Specification | LEDLM-80PL |
|---|---|
| Standards | IES LM-80-21 and IES TM-21-19 |
| Samples | Single LEDs or chips, R&D and quality testing |
| Life projection | Arrhenius model + TM-21-19 |
| Test fixtures | Two sets made for your LED type (more can be made) |
| Samples per test | Up to 16 (more on request) |
| Lumen maintenance criterion | L70, L50 or user-defined |
| Test options | Full 6,000-hour test or Arrhenius accelerated projection |
| Data recorded | Luminous flux, chromaticity and electrical parameters against time |
| Required configuration | |
| Power supply | DC12010 digital CC/CV DC power supply (sized to total sample power) |
| Temperature monitoring | TMP-8 eight-channel temperature tester |
| Test chamber | GDJW high/low temperature or GW high temperature chamber (sized to samples) |
LM-80 measures the lumen maintenance of LED packages, arrays and modules. LM-84 measures complete LED lamps and luminaires. The LEDLM-80PL is for LM-80; the LEDLM-84PL is for LM-84.
No. LISUN specifies a GDJW or GW series chamber, a DC12010 power supply and a TMP-8 temperature tester as required items, sized to your samples.