LISUN HLST thermal shock chambers, basket (two-zone) or air-switching (three-zone) type: −20/−40/−50 to 150°C, switching ≤10s, recovery ≤5min.
Enquire About This Product
The LISUN HLST series exposes samples to sudden changes between high and low temperature, to find failures caused by thermal expansion and contraction in solder joints, components, materials and assemblies. It covers −50°C to 150°C and suits laboratory testing of small samples as well as batch verification on a production line.
Separate hot and cold chambers are held at their set temperatures. The samples sit in a basket that a shuttle mechanism moves between the two chambers, so the sample goes from hot to cold, or cold to hot, in 10 seconds or less.
A hot chamber and a cold chamber are pre-heated and pre-cooled, and the samples stay still in a separate test chamber. Dampers switch the airflow so pre-cooled or pre-heated air floods the test chamber, which reaches the target temperature within the recovery time. Because the samples do not move, powered or cabled samples are easier to test, and the test chamber holds two shelves.
Refrigeration uses TECUMSEH or Bitzer hermetic compressors in a cascade system, and heating is by independent nichrome heaters. Small models are air cooled; larger models and the widest temperature ranges are water cooled. A colour touchscreen controls the chamber with 0.1°C resolution, and a Pt100 sensor measures temperature. Protection covers leakage, short circuit, heater over-temperature, motor overheating, and compressor over-pressure and over-current.
| Model | HLST-100D | HLST-250D | HLST-500D | HLST-100T | HLST-250T | HLST-500T |
|---|---|---|---|---|---|---|
| Type | Two-zone basket | Three-zone air switching | ||||
| Working space (mm) | 450 × 450 × 500 | 600 × 600 × 600 | 800 × 800 × 800 | 450 × 450 × 500 | 600 × 600 × 600 | 800 × 800 × 800 |
| Basket size (mm) | 250 × 250 × 250 | 400 × 400 × 450 | 600 × 600 × 600 | – | – | – |
| Sample shelves | 1 | 1 | 1 | 2 | 2 | 2 |
| Power, range A (−20 to 150°C) | ~15kW | ~25kW | ~38kW | ~20kW | ~30kW | ~40kW |
| Power, range B (−40 to 150°C) | ~22kW | ~32kW | ~45kW | ~28kW | ~38kW | ~45kW |
| Power, range C (−50 to 150°C) | ~32kW | ~43kW | ~48kW | ~35kW | ~43kW | ~48kW |
| Switching time | ≤10s | |||||
| Recovery time | ≤5min | |||||
| Temperature fluctuation / deviation / uniformity | ≤±0.5°C / ±2°C / ≤2°C | |||||
Add the temperature range letter A, B or C to the model. Cooling is by air or water depending on model and range.
The two-zone basket type gives the fastest transfer and is common for components. The three-zone type keeps samples still, which is better for powered samples, cabled samples or fragile parts.
The time for the air around the samples to reach the set temperature after the switch; the HLST does this in 5 minutes or less.