LISUN LSHF61000-3 single-phase harmonics and flicker test system to IEC 61000-3-2 and 61000-3-3: analyzer, flicker impedance and low-distortion AC source.
Enquire About This Product
Mains-powered products must keep the harmonic currents they draw, and the voltage flicker they cause, within the limits of IEC 61000-3-2 and IEC 61000-3-3. LISUN's harmonics and flicker test system is a complete single-phase setup for those tests, made of three matched instruments: the LSHF61000-3-2/3 harmonics and flicker analyzer, the LSAIF-503 flicker reference impedance and a LSP-HF low-distortion AC power source.
A high-speed 16-bit AD converter measures power quality in real time. The software assesses harmonics up to the 50th order and applies the limits for Class A, B, C and D equipment automatically, with a Pass/Fail result. For flicker it measures dc, dmax, Tmax, Pst and Plt over a configurable observation period and compares them with the IEC 61000-3-3 limits.
| Instrument | Specification |
|---|---|
| LSHF61000-3-2/3 harmonics analyzer | Voltage 1~600V (CF 1.67), ±0.05% of range ±0.05% of reading; current 2mA~13.3A (CF 3), ±0.05% of range ±0.05% of reading; frequency 45~65Hz, ±0.1%; power factor 0~1, ±0.005; Class 1 to IEC 61000-3-2 |
| LSHF61000-3-2/3 flicker analyzer | Flicker meter to IEC 61000-3-3, accuracy better than 5% |
| LSAIF-503 flicker impedance | Line 0.24Ω + j0.15Ω, neutral 0.16Ω + j0.10Ω, single phase 0.4Ω; to IEC 61000-3-3; max 16A |
| Low-distortion AC power source | LSP-500VAC-HF 500VA; LSP-1KVAC-HF 1kVA; LSP-2KVAC-HF 2kVA; LSP-3KVAC-HF 3kVA; LSP-5KVAC-HF 5kVA; all 0~300V |
The harmonic test only means something if the supply itself is clean. IEC 61000-3-2 sets tight limits on supply voltage distortion, which the LSP-HF sources are designed to meet.
Yes. This system covers single-phase equipment up to 16A.