LISUN GDJW-010A programmable temperature chamber without humidity: 1000 × 1000 × 1000mm, −20 to 150°C, ±0.5°C, for cold, dry heat and temperature cycling.
Enquire About This Product
The LISUN GDJW-010A is a programmable high and low temperature chamber with a one cubic metre working space. It runs cold, dry heat and temperature change tests (IEC 60068-2-1, 2-2 and 2-14 type tests) and low-temperature storage for materials and products, without humidity control. The A suffix denotes the −20 to 150°C range.
It suits electronic components and assemblies, LED products, automotive parts and materials where humidity is not needed, and shares its construction with LISUN's GDJS temperature and humidity chambers.
The inner chamber is SUS304 stainless steel, insulated with rigid polyurethane foam and fine glass fibre, with high-temperature silicone door seals. Temperature is measured by a Pt100 sensor. Refrigeration uses an air-cooled TECUMSEH compressor, heating is by independent nichrome heaters, and a low-noise centrifugal fan circulates the air evenly.
A LISUN controller with PLC runs programs and has USB, RS-232 and RS-485 ports for remote monitoring and data logging. Protection covers earth leakage, short circuit, heater and motor over-temperature, and compressor over-pressure, overload and over-current.
| Parameter | Value |
|---|---|
| Working space | 1000 × 1000 × 1000mm |
| Temperature range | −20 to 150°C (range A) |
| Fluctuation / uniformity | ±0.5°C / ±2°C |
| Heating rate | 1.0–3.0°C/min |
| Cooling rate | 0.7–1.0°C/min |
| Humidity | Not controlled |
| Supply | 380V ±10% three phase, 50/60Hz |
| Interfaces | USB, RS-232, RS-485 |
The GDJW series runs from 100L to 2000L, with B (−40°C), C (−60°C) and D (−70°C) ranges. For humidity testing, see the GDJS series; for heat only up to 250°C, the GW oven series.
No. If you need humidity, choose the GDJS version of the same size.
Yes. The LISUN GW series ovens go to 250°C (300°C option).