GLE-20 Mask Particulate Filtration Efficiency Tester – LISUN

GLE-20

LISUN GLE-20 oil and salt PFE tester for face masks and filter media: NaCl and oil aerosols, 10–100 L/min, filtration efficiency and airflow resistance.

Material & Optical Analysis
Other Analyzers
Bench-top Instrument
ASTM F2100; ASTM F2299; ASTM F1862; EN 14683; EN 149; EN 143; EN 140; EN 136; EN 14387; EN 1822-3; GB 2626; GB/T 32610; GB 19083; YY 0469; YY/T 0969
Enquire About This Product
LISUN GLE-20 Mask Particulate Filtration Efficiency Tester

Overview

The LISUN GLE-20 measures the particulate filtration efficiency (PFE) of face masks, respirators and flat filter media, and the resistance the sample presents to a constant airflow. It is used by mask manufacturers, textile and PPE test laboratories, medical device inspection centres and hospitals.

The GLE-20 is an oil-and-salt two-in-one system. It has two completely independent aerosol paths: a salt (NaCl) aerosol generator for KN-type filter elements and an oil aerosol generator (DOP, or equivalents such as DEHS or paraffin oil) for KP-type elements. Because the two hardware sets are separate, an oil test does not contaminate a salt test and vice versa.

How the test works

Aerosol is drawn through a 100 cm² test area of the sample at a set flow rate between 10 and 100 L/min. Particle concentration sensors measure upstream and downstream concentrations, and the system calculates filtration efficiency while a differential pressure sensor (0–1000 Pa) records breathing resistance. Salt particles have a count median diameter of 0.075 ±0.02 µm and oil particles 0.185 ±0.02 µm, as the respirator test standards specify.

Testing, data storage and reporting run from a touch-screen computer. Results can be exported, printed on A4 or on a small receipt printer, and reports can be printed in English.

Key Specifications

Specifications

ParameterValue
Flow range10–100 L/min, accuracy class 2.5
Filtration efficiency range0.001–99.999%
Sampling frequency1–9999 times/min, settable
Particle concentration range0.001–200 mg/m³
Differential pressure range0–1000 Pa
Salt particle CMD0.075 ±0.02 µm, geometric standard deviation ≤1.86
Oil particle CMD0.185 ±0.02 µm, geometric standard deviation ≤1.60
Aerosol concentrationNaCl 12–20 mg/m³; DOP 50–200 mg/m³
Test area100 cm²
AerosolsNaCl; DOP, DEHS, paraffin oil or corn oil optional
Ambient monitoringBuilt-in temperature and humidity sensor
Power supplyAC 220V 50Hz (120V 60Hz available)

Filtration efficiency classes (GB 2626)

ClassNaCl aerosolOil aerosol
KN90 / KN95 / KN10090.0% / 95.0% / 99.97%–
KP90 / KP95 / KP100–90.0% / 95.0% / 99.97%

Features

Applications

Frequently asked questions

Why are both salt and oil aerosols needed?

Filters rated for non-oily particles (KN, N, most FFP tests) are tested with NaCl, while filters rated for oily particles (KP, R, P) must also pass an oil aerosol test. The GLE-20 does both without cross-contamination.

Can it test flat material as well as finished masks?

Yes. The 100 cm² test area holds a flat sample of mask or filter material; finished masks can be tested with a suitable fixture.