LISUN
Home
/
EMC & EMI Test Equipment

ESD61000-2 ESD Simulator Gun (IEC 61000-4-2) – LISUN

ESD61000-2P / ESD61000-2 / ESD61000-2A / ESD61000-2C

LISUN ESD61000-2 ESD simulator guns to IEC/EN 61000-4-2 (ESD61000-2C also ISO 10605): 20kV or 30kV, contact and air discharge, Android touch screen.

EMC & EMI Test Equipment
ESD Simulator
Desktop / Portable Instrument
IEC 61000-4-2:2025; EN 61000-4-2:2023; ISO 10605 (ESD61000-2C); GB/T 17626.2-2018; GB/T 19951-2019 (ESD61000-2C)
Enquire About This Product
LISUN ESD61000-2 ESD Simulator Gun (IEC 61000-4-2)

Overview

An ESD simulator, or ESD gun, reproduces the static discharge from a person touching a product, so you can check that electronic equipment keeps working when it is zapped. LISUN's ESD61000-2 series generates the standard IEC 61000-4-2 discharge current waveform in contact and air discharge modes, with a 0.6 to 1ns rise time.

A large Android touch screen in English and Chinese holds built-in test routines for IEC/EN 61000-4-2 and GB/T 17626.2, recalled with one touch, plus single, counting, 20 pulses per second, programmed and custom modes. It is used for EMC compliance testing in LED lighting, consumer electronics, automotive electronics, industrial control and medical devices.

Choosing a model

Accessories

LISUN also supplies the ESD-DESK test bench with horizontal and vertical coupling planes, the ESD-CALA calibration target kit, the ESD-HV30KV1G high-impedance voltmeter for air discharge voltage checks, and the ESD-FCP field coupling plate for ISO 10605 Annex F.

Key Specifications

Specifications

SpecificationESD61000-2PESD61000-2ESD61000-2AESD61000-2C
StylePortableDesktop
RelayChinese brand, 200,000 operationsKilovac (USA), 1 million operations
Output voltage0.1~20kV ±5%0.1~30kV ±5%
PolarityPositive, negative, alternating
Discharge network150pF/330Ω (replaceable)150pF/330Ω, 150pF/2000Ω, 330pF/2000Ω, 330pF/330Ω
Current rise time0.6~1ns
Test modesSingle, counting, 20PPS, IEC level, programmed, custom; contact, air and interval discharge
TriggerManual or automatic
Discharge count / interval1~9,999 / 0.05~99.99s
ElectrodesPointed (contact) and round (air)Pointed, round and 30mm large round
SupplyAC 100~240V, 50/60Hz
StandardsIEC/EN 61000-4-2IEC/EN 61000-4-2 and ISO 10605

Calibration data (third-party CNAS report)

Output voltageFirst peak currentCurrent at 30nsCurrent at 60nsRise time
+2kV7.29A4.10A2.20A0.93ns
+4kV15.40A7.90A4.30A0.97ns
+6kV23.20A12.10A6.50A0.97ns
+8kV29.40A16.20A9.30A0.89ns
-8kV31.80A16.10A8.20A0.90ns

Features

Applications

Frequently asked questions

Which model do I need for automotive ESD testing?

The ESD61000-2C. It has the 150pF/330Ω, 150pF/2000Ω, 330pF/2000Ω and 330pF/330Ω networks that ISO 10605 calls for, and outputs up to 30kV.

Do I need a test table?

IEC 61000-4-2 specifies a ground reference plane and horizontal and vertical coupling planes. LISUN's ESD-DESK provides these.

How is the ESD gun calibrated?

LISUN offers the ESD-CALA target kit for current waveform calibration and the ESD-HV30KV1G meter for air discharge voltage, to IEC 61000-4-2.