LISUN EDX-2 energy dispersive XRF spectrometers for RoHS screening, alloy analysis (S–U, or Na–U on EDX-2T) and coating thickness, 2ppm to 99.99%.
Enquire About This Product
The LISUN EDX-2 series are energy dispersive X-ray fluorescence (EDXRF) spectrometers. An X-ray tube excites the sample, and each element in it emits X-rays at its own characteristic energy. The detector identifies the elements from those energies and works out how much of each is present from the intensity of each peak.
XRF is the standard fast screening method for RoHS. It needs little or no sample preparation and measures lead (Pb), mercury (Hg), cadmium (Cd), total chromium (Cr) and bromine (Br), the elements behind the six original RoHS restricted substances.
All models have a camera and chamber lighting so the operator can see exactly where on the sample the measurement is taken. LISUN supplies a silver reference and an EC681m reference material with calibration certificate.
| Parameter | EDX-2A / 2AC / 2AB / 2ABC | EDX-2T |
|---|---|---|
| Type | Desktop, in air | Desktop, vacuum |
| Detector | Si-PIN | SDD |
| Resolution | 149 eV | 129 eV |
| Tube voltage / current | 50kV / 600µA, set automatically | 50kV / 600µA, set automatically |
| Test time | 200 s | 100 s |
| Sample chamber | 610 × 320 × 100mm | 510 × 310 × 120mm (non-vacuum); Ø100 × 70mm (vacuum) |
| Sample types | Solid, liquid, powder | Solid, liquid, powder (non-vacuum); solid (vacuum) |
| Content range | 2ppm to 99.99% | 2ppm to 99.99% |
| Alloy element range | S (16) to U (92), EDX-2AB / 2ABC only | Na (11) to U (92) |
| Weight | 50kg | 55kg |
| Substance | Maximum concentration |
|---|---|
| Lead (Pb) | 1000ppm |
| Mercury (Hg) | 1000ppm |
| Cadmium (Cd) | 100ppm |
| Hexavalent chromium (Cr VI) | 1000ppm |
| PBB | 1000ppm |
| PBDE | 1000ppm |
XRF measures total Cr and Br. Samples that screen above the limits for these are confirmed for Cr VI, PBB and PBDE by wet chemistry.
XRF is a screening method. It measures total chromium and bromine, so samples near or over the limit are usually confirmed by chemical analysis for hexavalent chromium and the brominated flame retardants.
The EDX-2AB, EDX-2ABC or EDX-2T, which include alloy and precious metal analysis.