LISUN HTCT-300 CT and PT tester to IEC 60044-1 and 60044-6: excitation, ratio, polarity, resistance, burden, ALF and FS, knee points up to 30kV.
Enquire About This Product
The HTCT-300 tests protection, measurement and TP-class current transformers (CT) and electromagnetic voltage transformers (PT). It measures excitation (V-I) characteristics, ratio, polarity, secondary winding resistance, secondary burden, ratio error and phase error, and automatically reports the knee point voltage and current, 5% and 10% error curves, accuracy limit factor (ALF), instrument security factor (FS), secondary time constant, remanence factor and saturated and unsaturated inductance.
Using a low-frequency method, it can test CTs with knee points up to 30kV while staying portable at under 10kg. One wiring setup covers resistance, excitation, ratio and polarity in one click. It selects the IEC 60044-1 or 60044-6 test standard automatically for the transformer type and class, stores 2,000 results, and exports them by USB for reports in Word.
| Item | Specification |
|---|---|
| Supply | AC 220V ±10%, 50Hz |
| Output | 0~180Vrms, 12Arms, 36A peak |
| Test voltage accuracy | ±0.2% |
| CT ratio | 1~30,000, ±0.2% |
| PT ratio | 1~10,000, ±0.2% |
| Phase | 0.5 min resolution, ±5 min |
| Secondary winding resistance | 0~300Ω, 2% ±2mΩ |
| AC burden | 0~300VA, 2% ±0.2VA |
| Memory | 2,000 results, USB export |
| Environment | -10~40℃, ≤90% RH, ≤1000m altitude |
| Size / weight | 410×250×300mm / ≤10kg |
It uses a low-frequency test method, which reaches the core's saturation at a much lower voltage and then calculates the 50Hz equivalent.
IEC 60044-1 and IEC 60044-6 (and the equivalent GB standards), selected automatically for the transformer type.